Posted by John Kennard on October 31, 2006, 11:49 am
User logged in as UKAT5582
Hi just to advise anyone running these machines the problems we have experienced. The O2 cell keeps failing with a fault code logged. We have replaced in quick succession 2 new cells lasting about 3 weeks each. We have just had the Sun Engineer in who informed us that since the last software update that a bug exists and all you need to do is do a O2 re-install checking that the voltage reading for the cell is between 10-12.05 volts. Funnily enough our 3 week old cell reading 12.69 volts failed and had to be replaced under warranty, as the cell deteriorates the voltage should drop once below 10 volts it will not function. As if there is not enough hassle in life I hope this saves time for any one experiencing this problem.
|